The Calculation of SAR Distribution for a Water-cooled Microwave Dipole Applicator Radiating in a Biological Medium
watercooled microwave dipole applicator radiating in a biological tissue
is analyzed with a theoretical electromagnetic mode. The specific absorption rate (SAR)distribution is calculation is calculated taking into account the effect of the water flowing inside the applicator
and compaed with the SAR distribution of no-water-cooled radiating dipole. The result suggests that
the SAR distribution will be improved on the extension of the therapy depth due to the effect of water-cooled
and it comforms to the engineering experiments and clinical reality.