National Key Technology Research and Development Program New Generation of Information Technology (No.2014BAK02B03);Research on Time-Domain Transfer Impedance Measurement Technology of Public Welfare Industry in Quality Inspection (No.20140010)
ZHU Jiang-miao, ZHAO Lin-xiao, MIAO Jing-yuan, et al. Research on the Calibration Technology of Broadband Sampling Oscilloscope Based on the Standard Pulse Method of an Optical Pulse Source[J]. Acta Electronica Sinica, 2018, 46(4): 945-951.
DOI:
ZHU Jiang-miao, ZHAO Lin-xiao, MIAO Jing-yuan, et al. Research on the Calibration Technology of Broadband Sampling Oscilloscope Based on the Standard Pulse Method of an Optical Pulse Source[J]. Acta Electronica Sinica, 2018, 46(4): 945-951. DOI: 10.3969/j.issn.0372-2112.2018.04.024.
Research on the Calibration Technology of Broadband Sampling Oscilloscope Based on the Standard Pulse Method of an Optical Pulse Source
This paper constructed a calibration system of oscilloscope based on the optical pulse source. The 70GHz sampling module 86118A of Agilent 86100C was calibrated based on optical pulse source using the standard pulse method. In the calibration process
the influences on calibration results of three kinds of time base errors were researched for the first time
e. g.
time base drift was removed using a cross-correlation algorithm
time base distortion was corrected using the least squares method
and jitter was removed by using PDF deconvolution. Experiment results showed that calibration of 70GHz broadband oscilloscope was realized
the time base errors could be effectively removed by this method
and the accuracy of the calibration was satisfied.